Performance Evaluation of Spectroscopic Detectors for LEU Hold-up Measurements
Creators
- 1. Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States)
Description
The hold-up measurement of low-enriched uranium materials may require use of alternate detector types relative to the measurement of highly enriched uranium. This is in part due to the difference in process scale (i.e., the components are generally larger for low-enriched uranium systems), but also because the characteristic gamma-ray lines from 235U used for assay of highly enriched uranium will be present at a much reduced intensity (on a per gram of uranium basis) at lower enrichments. Researchers at Oak Ridge National Laboratory examined the performance of several standard detector types, e.g., NaI(Tl), LaBr3(Ce), and HPGe, to select a suitable candidate for measuring and quantifying low-enriched uranium hold-up in process pipes and equipment at the Portsmouth gaseous diffusion plant. Detector characteristics, such as energy resolution (full width at half maximum) and net peak count rates at gamma ray energies spanning a range of 60-1332 keV, were measured for the above-mentioned detector types using the same sources and in the same geometry. Uranium enrichment standards (Certified Reference Material no. 969 and Certified Reference Material no. 146) were measured using each of the detector candidates in the same geometry. The net count rates recorded by each detector at 186 keV and 1,001 keV were plotted as a function of enrichment (atom percentage). Background measurements were made in unshielded and shielded configurations under both ambient and elevated conditions of 238U activity. The highly enriched uranium hold-up measurement campaign at the Portsmouth plant was performed on process equipment that had been cleaned out. Therefore, in most cases, the thickness of the uranium deposits was less than the ''infinite thickness'' for the 186 keV gamma rays to be completely self-attenuated. Because of this, in addition to measuring the 186 keV gamma, the 1,001 keV gamma ray from 234mPa-a daughter of 238U in secular equilibrium with its parent-will also need to be measured. Based on the performance criteria of detection efficiency, energy resolution, peak-to-continuum ratios, minimum detectable limits, and the weight of the shielded probe, a shielded (0.5 in. thick lead shield) 2 x 2 in. NaI(Tl) detector is recommended for use. The recommended approach is to carry out analysis using data from both 186 keV and 1,001 keV gamma rays, and select a best result based on propagated uncertainty estimates. It is also highly recommended that a two-point gain stabilization scheme based on an 241Am seed embedded in the probe be implemented. Shielding configurations to reduce the impact of background interference on the measurement of 1,001 keV gamma-ray are discussed.
Availability note (English)
Available from http://info.ornl.gov/sites/publications/files/Pub71211.pdf; PURL: http://www.osti.gov/servlets/purl/1335357/Additional details
Identifiers
Publishing Information
- Imprint Pagination
- 44 p.
- Report number
- ORNL/TM--2016/649
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 48047421
- Subject category
- S11: NUCLEAR FUEL CYCLE AND FUEL MATERIALS; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Non-conventional Literature
- Descriptors DEI
- COUNTING RATES; ENERGY RESOLUTION; ENRICHED URANIUM; ENRICHMENT; EVALUATION; GAMMA RADIATION; HIGH-PURITY GE DETECTORS; KEV RANGE 100-1000; LANTHANUM BROMIDES; MEV RANGE 01-10; NAI DETECTORS; PERFORMANCE TESTING; PORTSMOUTH GASEOUS DIFFUSION PLANT; PROTACTINIUM 234; SHIELDS; SLIGHTLY ENRICHED URANIUM; THICKNESS; URANIUM 238
- Descriptors DEC
- ACTINIDE NUCLEI; ACTINIDES; ALPHA DECAY RADIOISOTOPES; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BROMIDES; BROMINE COMPOUNDS; DIMENSIONS; ELECTROMAGNETIC RADIATION; ELEMENTS; ENERGY RANGE; ENRICHED URANIUM; EVEN-EVEN NUCLEI; GASEOUS DIFFUSION PLANTS; GE SEMICONDUCTOR DETECTORS; HALIDES; HALOGEN COMPOUNDS; HEAVY NUCLEI; HOURS LIVING RADIOISOTOPES; INDUSTRIAL PLANTS; IONIZING RADIATIONS; ISOMERIC TRANSITION ISOTOPES; ISOTOPE ENRICHED MATERIALS; ISOTOPE SEPARATION PLANTS; ISOTOPES; KEV RANGE; LANTHANUM COMPOUNDS; LANTHANUM HALIDES; MATERIALS; MEASURING INSTRUMENTS; METALS; MEV RANGE; MINUTES LIVING RADIOISOTOPES; NATIONAL ORGANIZATIONS; NUCLEAR FACILITIES; NUCLEI; ODD-ODD NUCLEI; PROTACTINIUM ISOTOPES; RADIATION DETECTORS; RADIATIONS; RADIOISOTOPES; RARE EARTH COMPOUNDS; RESOLUTION; SCINTILLATION COUNTERS; SEMICONDUCTOR DETECTORS; SOLID SCINTILLATION DETECTORS; SPONTANEOUS FISSION RADIOISOTOPES; TESTING; URANIUM; URANIUM ISOTOPES; US DOE; US ERDA; US ORGANIZATIONS; YEARS LIVING RADIOISOTOPES
Optional Information
- Contract/Grant/Project number
- AC05-00OR22725
- Funding organization
- USDOE (United States)
- Secondary number(s)
- OSTIID--1335357