Influence of angular correlations on the full energy peak efficiency calibration using the sum peak method

  • 1. Faculty of Technical Sciences, Trg Dositeja Obradovića 6, 21000, Novi Sad (Serbia)
  • 2. Faculty of Sciences, Department of Physics, Trg Dositeja Obradovića 4, 21000, Novi Sad (Serbia)

Description

To investigate the accuracy of the calibration of the HPGe detectors using an uncalibrated complex gamma cascade emitting source and the sum-peak method, Monte Carlo simulations were conducted. It was shown that accuracy below 5% is achievable for a 133Ba source, only if the spectral data are corrected for angular correlations even at short distances from the detector endcap.