Published 2011 | Version v1
Journal article

Automatic computed tomography patient dose calculation using header metadata

  • 1. Centre de Recherche Public Henri Tudor, 29, Avenue John F. Kennedy, 1855 Luxembourg (Luxembourg)
  • 2. Hopital RHMS, Baudour (Belgium)
  • 3. Ministry of Health, Allee Marconi-Villa Louvigny, 2120 Luxembourg (Luxembourg)

Description

The present work describes a method that calculates the patient dose values in computed tomography (CT) based on metadata contained in DICOM images in support of patient dose studies. The DICOM metadata is pre-processed to extract necessary calculation parameters. Vendor-specific DICOM header information is harmonized using vendor translation tables and unavailable DICOM tags can be completed with a graphical user interface. CT-Expo, an MS Excel application for calculating the radiation dose, is used to calculate the patient doses. All relevant data and calculation results are stored for further analysis in a relational database. Final results are compiled by utilizing data mining tools. This solution was successfully used for the 2009 CT dose study in Luxembourg. National diagnostic reference levels for standard examinations were calculated based on each of the countries' hospitals. The benefits using this new automatic system saved time as well as resources during the data acquisition and the evaluation when compared with earlier questionnaire-based surveys. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncr338

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
147
Journal Issue
1-2
Journal Page Range
p. 317-320
ISSN
0144-8420

Conference

Title
International Conference on Radiation Protection in Medicine
Dates
1-3 Sep 2010
Place
Varna (Bulgaria)

Optional Information

Notes
6 refs