Published August 15, 1975
| Version v1
Journal article
Design for a conventional high-resolution neutron powder diffractometer
Description
The design considerations for a conventional neutron powder diffractometer are reconsidered. It is found that the resolution can be improved to the limits imposed by the powder particle size, while the effective intensity can at the same time be increased by using multiple counters with the correct combination of soller divergences, monochromator mosaic spread, and take-off angle. With the profile technique for data analysis, this high resolution diffractometer should permit the refinement of nuclear and magnetic structures having unit cells of up to 3500A3 volume. (Auth.)
Additional details
Identifiers
Publishing Information
- Journal Title
- Nuclear Instruments and Methods
- Journal Volume
- 127
- Journal Issue
- 3
- Series
- Nucl. Instrum. Methods.
- Journal Page Range
- 361-370
- ISSN
- 0029-554X
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 7222530
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S75: CONDENSED MATTER PHYSICS, SUPERCONDUCTIVITY AND SUPERFLUIDITY;
- Is Lead record
- Yes
- Descriptors DEI
- FOCUSING; NEUTRON DIFFRACTION; NEUTRON DIFFRACTOMETERS; PARTICLE SIZE; POWDERS; RESOLUTION; SCATTERING; SCATTERING AMPLITUDES; SPECIFICATIONS
- Descriptors DEC
- AMPLITUDES; COHERENT SCATTERING; DIFFRACTION; MEASURING INSTRUMENTS; SIZE