Published March 2003 | Version v1
Journal article

A two dimension position sensitive PPAC

Description

A two-dimensional positional sensitive parallel-plate avalanche detector (PPAC) with 100 mm x 100 mm active area has been developed for RIBLL. The detector consists of one anode and two cathodes. Each cathode position signal is readout with a resistance chain. The PPAC was tested at 700 Pa of flowing isobutane and +500 V at the anode, and an intrinsic position resolution of 0.76 mm (FWHM) was determined

Additional details

Additional titles

Augmented title (English)
The PPAC is stated for Parallel-Plate Avalanche Chamber

Publishing Information

Journal Title
Nuclear Techniques
Journal Volume
26
Journal Issue
3
Journal Page Range
p. 245-248
ISSN
0253-3219