The effect of field modifier blocks on the fast photoneutron dose equivalent from two high-energy medical linear accelerators
Creators
- 1. Dept. of Medical Physics, Tarbiat Modares Univ., PO Box 14115-151, Tehran (Iran, Islamic Republic of)
- 2. Gamma Irradiation Center, Atomic Energy Organization of Iran, End of Kargare Shomali, PO Box 11365-3486, Tehran (Iran, Islamic Republic of)
- 3. Dept. of Medical Physics and Engineering, Esfahan Univ. of Medical Sciences, Hezarjarib, Esfahan (Iran, Islamic Republic of)
- 4. Dept. of Radiology Technology, School of Paramedicine, Iran Univ. of Medical Sciences, Hemmat Highway, Tehran (Iran, Islamic Republic of)
- 5. National Radiation Protection Dept., Iranian Nuclear Regulatory Authority, PO Box 14155-4494, Tehran (Iran, Islamic Republic of)
Description
High-energy linear accelerators (linacs) have several advantages, including low skin doses and high dose rates at deep-seated tumours. But, at energies more than 8 MeV, photonuclear reactions produce neutron contamination around the therapeutic beam, which may induce secondary malignancies. In spite of improvements achieved in medical linac designs, many countries still use conventional (non-intensity-modulated radiotherapy) linacs. Hence, in these conventional machines, fitting the beam over the treatment volume may require using blocks. Therefore, the effect of these devices on neutron production of linacs needs to be studied. The aim of this study was to investigate the effect of field shaping blocks on photoneutron dose in the treatment plane for two high-energy medical linacs. Two medical linacs, a Saturn 43 (25 MeV) and an Elekta SL 75/25 (18 MeV), were studied. Polycarbonate (PC) films were used to measure the fluence of photoneutrons produced by these linacs. After electrochemical etching of the PC films, the neutron dose equivalent was calculated at the iso-centre and 50 cm away from the iso-centre. It was noted that by increasing the distance from the centre of the X-ray beam towards the periphery, the photoneutron dose equivalent decreases rapidly for both the open and blocked fields. Increasing the energy of the photons causes an increase in the amount of photoneutron dose equivalent. At 25 MeV photon energy, the lead blocks cause a meaningful increase in the dose equivalent of photoneutrons. In this research, a 30% increase was seen in neutron dose contribution to central axis dose at the iso-centre of a 25 MeV irregular field shaped by lead blocks. It is concluded that lead blocks must be considered as a source of photoneutron production when treating irregular fields with high-energy photons. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncm421Additional details
Identifiers
- DOI
- 10.1093/rpd/ncm421;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 128
- Journal Issue
- 3
- Journal Page Range
- p. 359-362
- ISSN
- 0144-8420
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 43120537
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S43: PARTICLE ACCELERATORS;
- Descriptors DEI
- CONTAMINATION; DOSE EQUIVALENTS; DOSE RATES; DOSES; FILMS; HUMAN POPULATIONS; LINEAR ACCELERATORS; NEOPLASMS; PHOTONEUTRONS; PHOTONUCLEAR REACTIONS; X RADIATION
- Descriptors DEC
- ACCELERATORS; BARYONS; DISEASES; ELECTROMAGNETIC RADIATION; ELEMENTARY PARTICLES; FERMIONS; HADRONS; IONIZING RADIATIONS; NEUTRONS; NUCLEAR REACTIONS; NUCLEONS; PHOTONUCLEONS; POPULATIONS; RADIATIONS
Optional Information
- Notes
- 11 refs