Published November 1996 | Version v1
Journal article

Measurement of low radiation dose by photon counting technique

  • 1. Calcutta Univ. (India). Dept. of Physics

Description

Photon counting technique has been employed in lyoluminescence (LL) based dosimetric measurements and found to be very successful in precise measurement of low radiation dose. In this technique photon is counted one by one. Mannose has been chosen as the most favoured dosimetric material due to its highest LL-yield among the tissue equivalent saccharides. Since minimising the error requires acquisition of a large number of counts, one would prefer to take no less than 50 mg of mannose for evaluation of the smallest useful dose. It has been found that the smallest useful dose is 0.5 Gy at an error of 3.5% while the threshold dose is 0.1 Gy. (author). 19 refs., 4 figs

Additional details

Publishing Information

Journal Title
Indian Journal of Pure and Applied Physics
Journal Volume
34
Journal Issue
11
Journal Page Range
p. 866-870.
ISSN
0019-5596
CODEN
IJOPAU