Development of the LBNL positron emission mammography camera
Description
We present the construction status of the LBNL Positron Emission Mammography (PEM) camera, which utilizes a PET detector module with depth of interaction measurement consisting of 64 LSO crystals (3x3x30 mm3) coupled on one end to a single photomultiplier tube (PMT) and on the opposite end to a 64 pixel array of silicon photodiodes (PDs). The PMT provides an accurate timing pulse, the PDs identify the crystal of interaction, the sum provides a total energy signal, and the PD/(PD+PMT) ratio determines the depth of interaction. We have completed construction of all 42 PEM detector modules. All data acquisition electronics have been completed, fully tested and loaded onto the gantry. We have demonstrated that all functions of the custom IC work using the production rigid-flex boards and data acquisition system. Preliminary detector module characterization and coincidence data have been taken using the production system, including initial images
Availability note (English)
Available from INIS in electronic form; Also available from OSTI as DE00808923; PURL: https://www.osti.gov/servlets/purl/808923-5iuHaF/native/
Files
Additional details
Identifiers
Publishing Information
- Imprint Pagination
- 6 p.
- Report number
- LBNL--50791
Conference
- Title
- IEEE Nuclear Science Symposium and Medical Imaging Conference
- Dates
- 10-16 Nov 2002
- Place
- Norfolk, VA (United States)
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 34030806
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CAMERAS; CONSTRUCTION; DATA ACQUISITION; DATA ACQUISITION SYSTEMS; PHOTODIODES; PHOTOMULTIPLIERS; POSITRON COMPUTED TOMOGRAPHY; POSITRONS; SILICATES
- Descriptors DEC
- ANTILEPTONS; ANTIMATTER; ANTIPARTICLES; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; ELEMENTARY PARTICLES; EMISSION COMPUTED TOMOGRAPHY; FERMIONS; LEPTONS; MATTER; OXYGEN COMPOUNDS; PHOTOTUBES; SEMICONDUCTOR DEVICES; SEMICONDUCTOR DIODES; SILICON COMPOUNDS; TOMOGRAPHY
Optional Information
- Contract/Grant/Project number
- B and R KP1401030; AC03-76SF00098
- Funding organization
- USDOE Director, Office of Science. Office of Biological and Environmental Research. Medical Sciences Division (United States); National Institutes of Health (United States)