Published July 1984
| Version v1
Journal article
Performance characteristics of A150 plastic-equivalent gases in A150 plastic proportional counters for 14.8-MeV neutrons
- 1. Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin 53706
Description
Two recently developed A150 plastic-equivalent gas mixtures have been tested for suitability in proportional counter applications. Methane- and propane-based ''tissue-equivalent'' gases were also included for comparison purposes. Event-size weighted dose distributions were measured in a 14.8-MeV neutron beam. Resolution was found to be independent of gas mixture. Moreover the gains of the two A150 mixtures were the same, and comparable to that of the methane-based gas mixture. The ionization yield per event size was also independent of the hydrogenous gas mixture employed. Neutron doses determined with the proportional counter were in reasonable agreement with those obtained from an ionization chamber
Additional details
Publishing Information
- Journal Title
- Med. Phys.
- Journal Volume
- 11
- Journal Issue
- 4
- Series
- Med. Phys.
- Journal Page Range
- 449-455
- ISSN
- 0094-2405
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 16011133
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- IONIZATION; IONIZATION CHAMBERS; METHANE; MEV RANGE 10-100; NEUTRON BEAMS; NEUTRON DOSIMETRY; PERFORMANCE; PROPANE; RADIATION DOSE DISTRIBUTIONS; RESOLUTION; TISSUE-EQUIVALENT DETECTORS; TISSUE-EQUIVALENT MATERIALS
- Descriptors DEC
- ALKANES; BEAMS; DOSIMETRY; ENERGY RANGE; HYDROCARBONS; MATERIALS; MEASURING INSTRUMENTS; MEV RANGE; NUCLEON BEAMS; ORGANIC COMPOUNDS; PARTICLE BEAMS; RADIATION DETECTORS