Compact cold stage for micro-computerized tomography imaging of chilled or frozen samples
- 1. Department of Land, Air, and Water Resources, University of California, Davis, California 95616 (United States)
- 2. Paige Instruments, Woodland, California 95776 (United States)
- 3. Center for Molecular and Genomic Imaging, Department of Biomedical Engineering, University of California, Davis, California 95616 (United States)
Description
High resolution X-ray microCT (computerized tomography) can be used to image a variety of objects, including temperature-sensitive materials. In cases where the sample must be chilled or frozen to maintain sample integrity, either the microCT machine itself must be placed in a refrigerated chamber, or a relatively expensive commercial cold stage must be purchased. We describe here the design and construction of a low-cost custom cold stage suitable for use in a microCT imaging system. Our device uses a boron nitride sample holder, two-stage Peltier cooler, fan-cooled heat sink, and electronic controller to maintain sample temperatures as low as −25 °C ± 0.2 °C for the duration of a tomography acquisition. The design does not require modification to the microCT machine, and is easily installed and removed. Our custom cold stage represents a cost-effective solution for refrigerating CT samples for imaging, and is especially useful for shared equipment or machines unsuitable for cold room use
Additional details
Identifiers
- DOI
- 10.1063/1.4871473;
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 85
- Journal Issue
- 4
- Journal Page Range
- p. 043708-043708.5
- ISSN
- 0034-6748
- CODEN
- RSINAK
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 45076097
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BLOWERS; BORON NITRIDES; COMPUTERIZED TOMOGRAPHY; CONSTRUCTION; EQUIPMENT; HEAT EXCHANGERS; HEAT SINKS; IMAGES; SAMPLE HOLDERS
- Descriptors DEC
- BORON COMPOUNDS; DIAGNOSTIC TECHNIQUES; NITRIDES; NITROGEN COMPOUNDS; PNICTIDES; SINKS; TOMOGRAPHY
Optional Information
- Notes
- (c) 2014 AIP Publishing LLC