Published March 1994 | Version v1
Journal article

Two-dimensional restoration of single photon emission computed tomography images using the Kalman filter

  • 1. Univ. of Calgary, Alberta (Canada). Dept. of Electrical and Computer Engineering
  • 2. Foothills Hospital, Calgary, Alberta (Canada). Div. of Nuclear Medicine

Description

The discrete filtered backprojection (DFBP) algorithm used for the reconstruction of single photon emission computed tomography (SPECT) images affects image quality because of the operations of filtering and discretization. The discretization of the filtered backprojection process can cause the modulation transfer function (MTF) of the SPECT imaging system to be anisotropic and nonstationary, especially near the edges of the camera's field of view. The use of shift-invariant restoration techniques fails to restore large images because these techniques do not account for such variations in the MTF. This study presents the application of a two-dimensional (2-D) shift-variant Kalman filter for post-reconstruction restoration of SPECT slices. This filter was applied to SPECT images of a hollow cylinder phantom; a resolution phantom; and a large, truncated cone phantom containing two types of cold spots, a sphere, and a triangular prism. The images were acquired on an ADAC GENESYS camera. A comparison was performed between results obtained by the Kalman filter and those obtained by shift-invariant filters. Quantitative analysis of the restored images performed through measurement of root mean squared errors shows a considerable reduction in error of Kalman-filtered images over images restored using shift-invariant methods

Additional details

Publishing Information

Journal Title
IEEE Transactions on Medical Imaging
Journal Volume
13
Journal Issue
1
Journal Page Range
p. 102-109.
ISSN
0278-0062
CODEN
ITMID4

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
26001425
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ACCURACY; IMAGE PROCESSING; PERFORMANCE; QUALITY CONTROL; SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY
Descriptors DEC
COMPUTERIZED TOMOGRAPHY; CONTROL; EMISSION COMPUTED TOMOGRAPHY; TOMOGRAPHY