Published June 1, 2019 | Version v1
Journal article

Optimization of couch angles and number of arcs in non-coplanar VMAT for pituitary adenomas

  • 1. Master of Science Program in Medical Physics, Faculty of Medicine Ramathibodi Hospital, Mahidol University, Bangkok 10400 (Thailand)
  • 2. Radiosurgery Center, Faculty of Medicine Ramathibodi Hospital, Mahidol University, Bangkok 10400 (Thailand)
  • 3. Department of Diagnostic and Therapeutic Radiology, Faculty of Medicine Ramathibodi Hospital, Mahidol University, Bangkok 10400 (Thailand)

Description

Non-coplanar VMAT spares critical organs with short delivery time while maintaining PTV coverage. Due to the growth patterns of pituitary adenoma, the beam angle selection step was simplified by finding the optimal couch angles and number of arcs. The 2 clinical cases with difference PTV sizes were selected to study the results of dose distribution between coplanar and non-coplanar techniques. Later, couch angles were varied from 30° to 90° and 270° to 330° with a 15° interval. The numbers of arcs were varied from 3 to 4 arcs. Then, the selected protocol was applied to 5 pituitary adenoma patients. It was found that 90° and 270° couch angle showed the highest scores. However, these beams passed directly through the body, so the second couch angles of 75° and 285° were selected instead. For number of arcs optimization, 3-arc plan and 4-arc plan demonstrated a similar score. However, the 3-arc plan illustrated an obviously shorter treatment time. After the selection was proven to the 5 patients, there were 4 patients who passed the criteria, while 1 patient did not pass the hippocampal planning criteria. In conclusion, the suggested couch angles were 75° and 285°. The optimal number of arcs for pituitary adenoma was 3 arcs. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/1248/1/012059

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
1248
Journal Issue
1
Journal Page Range
[6 p.]
ISSN
1742-6596

Conference

Title
18. Asia-Oceania Congress of Medical Physics (AOCMP); 16. South-East Asia Congress of Medical Physics (SEACOMP)
Dates
11-14 Nov 2018
Place
Kuala Lumpur (Malaysia)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
53055645
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ADENOMAS; BEAMS; CRITICAL ORGANS; OPTIMIZATION; PATIENTS; RADIATION DOSE DISTRIBUTIONS; RADIOTHERAPY
Descriptors DEC
BODY; CARCINOMAS; DISEASES; MEDICINE; NEOPLASMS; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; THERAPY