Microwave Ablation in the Proximity of Surgical Clips: Is there a Safety Issue?
Creators
- 1. Hôpital Kirchberg (Hôpitaux Robert Schuman) (Luxembourg)
- 2. University Hospital RWTH Aachen. Department of Diagnostic and Interventional Radiology (Germany)
Description
Purpose
: The purpose of this study was to evaluate the heat generation of surgical clips within the target area of MWA and the influences on the ablation volume.Materials and Methods
: In bovine liver tissue, 42 ex vivo microwave ablations (60 W; 180 s) were performed. During ablation, the temperature was measured continuously at 4 points of interest (POI), in a distance of 7.5 and 15 mm on each side of the microwave antenna, with a titanium surgical placed at one 7.5-mm POI. Ablation volumes containing large vessels (n = 10) were excluded. For every POI, the mean temperature of 32 ablations was calculated. The mean temperatures were compared between the 4 POI and statistically analyzed using the Student's t test.Results
: The mean maximum temperatures at the side of the clip were 88.76 °C/ 195 s and 52.97 °C/ 195 s and at the side without clip 78.75 °C/ 195 s and 43.16 °C/ 195 s, respectively, at POI 7.5 mm and POI 15 mm. The maximum difference of mean temperatures for POI 7.5 mm was 12.91 °C at 84 s (p = 0.022) and for POI 15 mm 9.77 °C at 195 s (p = 0.009). No significant changes in size and shape of the ablation zone could be determined.Conclusions
: Our study demonstrated significantly higher temperatures adjacent to surgical clips. Also, the temperatures distal to the titanium clip were higher compared to the control location without clip. These findings suggest an increased risk of thermal damage to surrounding tissues during MWA, especially in case of immediate contact to surgical clips.Additional details
Identifiers
Publishing Information
- Journal Title
- Cardiovascular and Interventional Radiology
- Journal Volume
- 43
- Journal Issue
- 6
- Journal Page Range
- p. 918-923
- ISSN
- 0174-1551
- CODEN
- CAIRDG
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 55061255
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ABLATION; ANIMAL TISSUES; ANTENNAS; CATTLE; COMPARATIVE EVALUATIONS; CONTROL; DAMAGE; HAZARDS; LIVER; MICROWAVE RADIATION; SAFETY; SHAPE; SURGERY; TITANIUM; VOLUME
- Descriptors DEC
- ANIMALS; BODY; DIGESTIVE SYSTEM; DOMESTIC ANIMALS; ELECTRICAL EQUIPMENT; ELECTROMAGNETIC RADIATION; ELEMENTS; EQUIPMENT; EVALUATION; GLANDS; MAMMALS; MEDICINE; METALS; ORGANS; RADIATIONS; RUMINANTS; TRANSITION ELEMENTS; VERTEBRATES
Optional Information
- Copyright
- Copyright (c) 2020 © The Author(s) 2020