Published May 1976 | Version v1
Journal article

Measurement of primary bremsstrahlung spectrum from an 8-MeV linear acceleration

  • 1. Department of Radiology, Brooke Army Medical Center, Fort Sam Houston, Texas 78234

Description

The bremsstrahlung spectrum from an 8-MeV linear accelerator was measured using a NaI(T1) spectrometer system. The spectrum shows a low-energy cutoff at 0.4 MeV and the maximum photon energy to be approximately 6% greater than the nominal energy. The maximum emission of energy fluence was 1.6 and 1.8 MeV for measured and calculated values, respectively. The fast neutron dose in the photon beam was approximately 0.09% of the x-ray dose. The weighted mean energy was 2.3 MeV, measured value, and 2.4 MeV, calculated value