Published 2002 | Version v1
Miscellaneous

A comparison study for dose calculation in radiation therapy: pencil beam Kernel based vs. Monte Carlo simulation vs. measurements

  • 1. The Catholic Univ. of Korea, Seoul (Korea, Republic of)
  • 2. Kangnam St. Mary's Hospital, Seoul (Korea, Republic of)

Description

Accurate dose calculation in radiation treatment planning is most important for successful treatment. Since human body is composed of various materials and not an ideal shape, it is not easy to calculate the accurate effective dose in the patients. Many methods have been proposed to solve inhomogeneity and surface contour problems. Monte Carlo simulations are regarded as the most accurate method, but it is not appropriate for routine planning because it takes so much time. Pencil beam kernel based convolution/superposition methods were also proposed to correct those effects. Nowadays, many commercial treatment planning systems have adopted this algorithm as a dose calculation engine. The purpose of this study is to verify the accuracy of the dose calculated from pencil beam kernel based treatment planning system comparing to Monte Carlo simulations and measurements especially in inhomogeneous region. Home-made inhomogeneous phantom, Helax-TMS ver. 6.0 and Monte Carlo code BEAMnrc and DOSXYZnrc were used in this study. In homogeneous media, the accuracy was acceptable but in inhomogeneous media, the errors were more significant. However in general clinical situation, pencil beam kernel based convolution algorithm is thought to be a valuable tool to calculate the dose

Additional details

Publishing Information

Publisher
KARP
Imprint Place
Taejon (Korea, Republic of)
Imprint Title
Proceedings of the first Asian and Oceanic Congress for Radiation Protection(AOCRP-1)
Imprint Pagination
[1 CD-ROM]
Journal Page Range
[5 p.]

Conference

Title
1. Asian and Oceanic Congress for Radiation Protection(AOCRP-1)
Dates
20-24 Oct 2002
Place
Seoul (Korea, Republic of)

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
36054031
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Quality check status
Yes
Descriptors DEI
ALGORITHMS; B CODES; COMPUTER CALCULATIONS; KERNELS; MONTE CARLO METHOD; PATIENTS; RADIATION DOSES; RADIOTHERAPY; SIMULATION;
Descriptors DEC
CALCULATION METHODS; COMPUTER CODES; DOSES; MATHEMATICAL LOGIC; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY;

Optional Information

Lead record
p5yyy-fem83
Notes
8 refs, 3 figs