A Simple Technique of Supine Craniospinal Irradiation
Creators
- 1. Department of Radiation Oncology, Tata Memorial Hospital, Mumbai, Maharashtra (India)
Description
We describe a simple procedure of craniospinal irradiation in supine position. The procedure was carried out with a 100-cm isocenter linear accelerator and compatible simulator. Treatment was with a 1 or 2 posteroanterior (PA)-directed spinal fields abutting lateral-directed cranial fields. Abutment of the fields was established by placement of markers on the neck of the patient, which provided a measure of the divergence of the spinal field. The precision and reproducibility of this technique, including the placement of junctions, appeared to be as good as for treatment in the prone position. The same could be verified with port films. We conclude that this new technique of supine craniospinal treatment is a simple and convenient alternative to traditional treatment in the prone position
Availability note (English)
Available from http://dx.doi.org/10.1016/j.meddos.2007.03.004Additional details
Identifiers
- DOI
- 10.1016/j.meddos.2007.03.004;
- PII
- S0958-3947(07)00056-8;
Publishing Information
- Journal Title
- Medical Dosimetry
- Journal Volume
- 33
- Journal Issue
- 1
- Journal Page Range
- p. 1-5
- ISSN
- 0958-3947
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 39064669
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; FILM DOSIMETRY; IRRADIATION; LINEAR ACCELERATORS; NECK; PHOTOGRAPHIC FILM DOSEMETERS; SIMULATORS
- Descriptors DEC
- ACCELERATORS; ANALOG SYSTEMS; BODY; DOSEMETERS; DOSIMETRY; FUNCTIONAL MODELS; MEASURING INSTRUMENTS