Published 2021 | Version v1
Journal article

Update of the BIPM comparison BIPM.RI(II)-K1.Gd-153 of activity measurements of the radionuclide 153Gd to include the 2020 result of the NIST (United States)

  • 1. Bureau International des Poids et Mesures, Pavillon de Breteuil, F-92312 Sevres Cedex, (France)
  • 2. National Institute of Standards and Technology (NIST), 100 Bureau Drive, Mail Stop 8462 MD 20899-8462 Gaithersburg, (United States)

Description

Since 1988, 3 laboratories have submitted 5 samples of 153Gd to the International Reference System (SIR) for activity comparison at the Bureau International des Poids et Mesures (BIPM), with comparison identifier BIPM.RI(II)- K1.Gd-153. Recently, the NIST (United States) participated in the comparison and the key comparison reference value (KCRV) has been updated. The degrees of equivalence between each equivalent activity measured in the SIR and the updated KCRV have been calculated and the results are given in the form of a table. A graphical presentation is also given. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1088/0026-1394/58/1A/06027

Additional details

Publishing Information

Journal Title
Metrologia
Journal Volume
58
Journal Issue
no.1A
Journal Page Range
p. 06027.1-06027.15
ISSN
0026-1394

Optional Information

Notes
11 refs.