Published January 2019 | Version v1
Journal article

Results of the CCRI(II)-S12.H-3 supplementary comparison: Comparison of methods for the calculation of the activity and standard uncertainty of a tritiated-water source measured using the LSC-TDCR method

  • 1. DRT, CEA LIST Laboratoire National Henri Becquerel (LNE-LNHB), CEA-Saclay, 91191 Gif sur Yvette Cedex, (France)
  • 2. European Commission, Joint Research Centre, (JRC), Directorate for Nuclear Safety and Security, Retieseweg 111, B-2440 Geel, (Belgium)
  • 3. Horia Hulubei National Institute for R and D in Physics and Nuclear Engineering (IFIN-HH), Magurele, RO 0-77125, (Romania)

Description

A comparison of calculations of the activity of a 3H2O liquid scintillation source using the same experimental data set collected at the LNE-LNHB with a triple-to-double coincidence ratio (TDCR) counter was completed. A total of 17 laboratories calculated the activity and standard uncertainty of the LS source, using the files with experimental data. The results, as well as relevant information on the computation techniques are presented and analysed in this paper. All results are compatible, even if there is a significant dispersion between the reported uncertainties. An output of this comparison is the estimation of the dispersion of TDCR measurement results when measurement conditions are well defined. (authors)

Availability note (English)

Available from doi: https://doi.org/10.1088/0026-1394/56/1A/06005

Additional details

Publishing Information

Journal Title
Metrologia
Journal Volume
56
Journal Issue
no.1A
Journal Page Range
p. 06005.1-06005.25
ISSN
0026-1394

Optional Information

Notes
21 refs.