Published 1987 | Version v1
Journal article

Radioactivity measurement of 3H-labeled gas with position-sensitive proportional counter

  • 1. Nagoya Univ. (Japan). Faculty of Engineering

Description

A position-sensitive proportional counter with a partially resistive anode wire was constructed for activity measurement of 3H-labeled methane and its characteristics were measured. The new method was able to completely eliminate the influence of the end effect of the counter and the effective volume of the counter was accurately measured with estimated uncertainty of +- 0.12%. The activity per unit volume 3H-labeled methane in the counter was determined with a tentatively estimated uncertainty of +-0.79% (at a level of one standard deviation). (author)

Additional details

Additional titles

Subtitle (English)
Pt.2. Characteristics of the counter and activity measurement

Publishing Information

Journal Title
Appl. Radiat. Isot.
Journal Volume
38
Journal Issue
5
Series
Appl. Radiat. Isot.
Journal Page Range
379-383
CODEN
ARISE