Published May 2009 | Version v1
Journal article

Research on the Effect of High Humidity on the Measuring of Surface α Activity

  • 1. Naval Medical Research Institute, Shanghai (China)

Description

This paper elaborates on the causes of errors in the measuring of surface α activity in highly humid environment by analyzing the effect of relative humidity on air density and surface adsorption. With the use of Φ45mm α source, surface activity response factors of MicrocontII surface α β measuring instrument are measured in 5 different relative humidity conditions at 20 degree C, thus basically showing the law of relative humidity affects the measuring of surface α activity, which can provide a reference for measuring surface α activity in highly humid conditions. (authors)

Additional details

Publishing Information

Journal Title
Nuclear Electronics and Detection Technology
Journal Volume
29
Journal Issue
3
Journal Page Range
p. 658-660, 709
ISSN
0258-0934

Optional Information

Notes
3 figs., 1 tab., 3 refs.