Simulated and associated experimental results of CdZnTe radiation detector response for gamma-ray imaging applications

  • 1. LETI-CEA, Grenoble (France)

Description

Simulated and associated experimental results of a high efficiency CdZnTe (CZT) radiation detector response for gamma-ray imaging applications are presented. The model of a high efficiency semiconductor gamma ray detector takes into account several different physical phenomena involved in the detection and correction processes, namely the geometry of the irradiation, the gamma-ray's interaction with the crystal, the physics of the semiconductor's charge collection, the electric field distribution and the pulse height correction method. A few important decoupling assumptions allow the authors to use a one dimensional charge collection simulation with a two-dimensional field model and a full three dimensional Monte-Carlo calculation of the gamma ray interactions. The model allows calculation of charge collection and gamma ray spectra for non uniform electric field distribution in either planar, striped or pixellated detector. The model takes also into account the new CZT fast pulse correction method and its associated noise by considering the pulse height and the rise time of electron signals (Bi-Parametric spectrum) for all gamma ray interactions. Specific simulated and experimental spectra at 122 keV are presented for CZT. First, basic spectral changes are calculated for variations in crystal and detector properties like mobility, trapping lifetime and electric field profiles. Second, new experimental results of the fast pulse correction method applied to different CZT detector grades are presented. This method allows to achieve a high detection efficiency (> 80%) with a good energy resolution (< 6% FWHM) at 122 keV for a 4 x 4 x 6 mm3 CZT detector. No specific contact geometry is needed and the unusual low applied bias voltage allows to limit the aging and break voltage effects and also the dark current and its associated noise. This fast correction method is expected to be useful for medical imaging and other applications. Finally, simulated Bi-Parametric (BP) spectra expected with the fast pulse correction method according to the detector properties (electric field profiles, electron lifetime) are simulated and a qualitative comparison is provided

Additional details

Publishing Information

Publisher
Materials Research Society
Imprint Place
Warrendale, PA (United States)
ISBN
1-55899-392-4
Imprint Title
Semiconductors for room-temperature radiation detector applications 2
Imprint Pagination
681 p.
Series
Materials Research Society symposium proceedings, Volume 487
Journal Page Range
p. 171-185
ISSN
0272-9172

Conference

Title
1997 fall meeting of the Materials Research Society
Dates
1-5 Dec 1997
Place
Boston, MA (United States)

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
30034734
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference, Numerical Data
Descriptors DEI
CADMIUM TELLURIDES; EXPERIMENTAL DATA; GAMMA SPECTROMETERS; MATHEMATICAL MODELS; RESPONSE FUNCTIONS; SEMICONDUCTOR DETECTORS; ZINC TELLURIDES
Descriptors DEC
CADMIUM COMPOUNDS; CHALCOGENIDES; DATA; FUNCTIONS; INFORMATION; MEASURING INSTRUMENTS; NUMERICAL DATA; RADIATION DETECTORS; SPECTROMETERS; TELLURIDES; TELLURIUM COMPOUNDS; ZINC COMPOUNDS

Optional Information

Secondary number(s)
CONF-971201--