Published March 2005 | Version v1
Book

A flow through ionization chamber detector for measuring week beta radiation in a gaseous stream

  • 1. Radiopharmaceuticals Division, Bhabha Atomic Research Centre, Mumbai (India)
  • 2. Radiochemistry and Isotope Group, Bhabha Atomic Research Centre, Mumbai (India)

Description

A flow through miniature ionization chamber of 1 mL active volume has been set up and calibrated for the measurement of tritium and other tritium bearing components emerging from the gas chromatographic column. Current -voltage characteristics for the chamber and Calibration curve have been worked out to measure tritium bearing components in flowing stream at a flow rate of 60 mL/min. With the help of this ionization chamber detector, it is possible to find tritium concentrations as low as 10 pCi/mL with an accuracy of + 5% RSD. (author)

Part of:
Proceedings of DAE-BRNS national symposium on compact nuclear instruments and radiation detector-2005

Additional details

Publishing Information

Publisher
Bhabha Atomic Research Centre
Imprint Place
Mumbai (India)
ISBN
81-88513-15-6
Imprint Title
Proceedings of DAE-BRNS national symposium on compact nuclear instruments and radiation detector-2005
Imprint Pagination
730 p.
Journal Page Range
p. 3-6

Conference

Title
compact nuclear instruments and radiation detectors-2005
Acronym
CNIRD-2005
Dates
2-4 Mar 2005
Place
Jodhpur (India)

Optional Information

Notes
3 refs., 5 figs.