Published 1993 | Version v1
Journal article

Beta camera low activity tumor imaging

  • 1. Dept. of Radiation Physics, Lund Univ. Hospital (Sweden)
  • 2. Dept. of Clinical Neurophysiology, Lund Univ. Hospital (Sweden)
  • 3. Dept. of Neuropathology, Lund Univ. Hospital (Sweden)
  • 4. Dept. of Neurosurgery, Lund Univ. Hospital (Sweden)

Description

A new technique, the beta camera, to complement film autoradiography, with fast quantitative imaging of beta particle-emitting radionuclides has been developed. It consists of a thin plastic scintillator and a light-sensitive microchannel plate detector. The thin tissue sample is mounted on the scintillator. Our first system had a high background and a moderate spatial resolution of 900 μm. We now report an improved system with a photomultiplier tube mounted on the scintillator of the microchannel plate detector. Only events registered by both detectors are accepted. A fast coincidence unit processes the signals, and if a time overlap exists, an event is generated in the beta camera. In the coincidence mode, images with low activity distribution of 201Tl (count rate 1 s-1) in 50 μm-thick slices of a human glioma tumor could be recorded with a spatial resolution of 500 μm. (orig.)

Additional details

Publishing Information

Journal Title
Acta Oncologica (Stockholm)
Journal Volume
32
Journal Issue
7-8
Journal Page Range
p. 869-872.
ISSN
0284-186X
CODEN
ACTOEL

Conference

Title
3. Scandinavian symposium on radiolabeled monoclonal antibodies for cancer diagnosis and therapy.
Dates
30-31 Oct 1992.
Place
Helsinki (Finland).