Published September 1, 2008 | Version v1
Journal article

Energy calibration of gamma spectra in plastic scintillators using Compton kinematics

  • 1. Pacific Northwest National Laboratory, MS K7-36, National Security Division, 1005 Country Court, P. O. Box 999, Richland, WA 99352 (United States)

Description

This paper describes a simple and practicable method for assigning energy values to gamma-ray pulse-height distributions measured with polyvinyl toluene (PVT) based detectors. It is based upon the characteristic shape of simulated spectra in the region of maximum energy deposition resulting from a single Compton scattering. The validity of this method is demonstrated by applying it to a set of measured NaI(Tl) spectra, and comparing those results to the standard photopeak method of calibrating the same spectra. The method is then applied to a set of measured PVT spectra. It can now be applied to calibration of fielded PVT detectors using two selected gamma-ray sources, which was the initial motivation for this study

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2008.06.031

Additional details

Identifiers

DOI
10.1016/j.nima.2008.06.031;
PII
S0168-9002(08)00877-2;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
594
Journal Issue
2
Journal Page Range
p. 232-243
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2008 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.