Full-text:Biodistribution (BQC) of 99mTc-DTPA in Sprague Dawley rats was evaluated in six (6) different time points within the range of 5 min to 24 h.Through cut and count approach, three (3) rats were used for each time point.The gamma radiation emitted from 99mTc isotope in 13 organs/samples was counted using Gamma Counter (Wizard2 Perkin Elmer, USA). Prior to use, the Gamma Counter shall undergo normalization, background measurement and Instrument Performance Assessment Measurement. Correction to the Gamma Counter efficiency was performed by applying radioisotope (99mTc) standard data to BQC data. This correction needs to be performed for every isotope used in BQC cut and count study. (author)

Availability note (English)

Available in abstract form only, full text entered in this record

Additional details

Publishing Information

Imprint Pagination
1 p.

Conference

Title
Nuclear Innovation and Technical Convention 2023
Acronym
NITC 2023
Dates
24-26 Oct 2023
Place
Bangi (Malaysia)

Optional Information

Notes
Poster presentation