Validation of gamma-ray true coincidence summing effects modeled by the Monte Carlo code MCNP-CP
- 1. Canberra Industries, Meriden, CT (United States)
- 2. National Academy of Science, Kiev (Ukraine). Institute for Nuclear Research
Description
When radionuclides decay by cascading photons, the accuracy of the measured nuclide activity may be affected by true coincidence summing effects. The effects can be quantified by Monte Carlo simulations that can handle correlated γ- and X-ray emissions from a radionuclide. Analysis techniques are also available commercially to correct for the effects due to cascading γ-rays. The MCNP-CP code was used to compute the effects in high purity germanium detectors for several commonly used nuclides and geometries and the results were compared to measurements and an analysis technique. Excellent agreement in true coincidence summing corrections predicted by MCNP-CP and the analysis technique was obtained. In addition, the X-ray true coincidence summing effects were evaluated. (author)
Additional details
Publishing Information
- Journal Title
- Journal of Radioanalytical and Nuclear Chemistry
- Journal Volume
- 278
- Journal Issue
- 2
- Journal Page Range
- p. 359-363
- ISSN
- 0236-5731
- CODEN
- JRNCDM
INIS
- Country of Publication
- Hungary
- Country of Input or Organization
- Hungary
- INIS RN
- 40031884
- Subject category
- S38: RADIATION CHEMISTRY, RADIOCHEMISTRY AND NUCLEAR CHEMISTRY;
- Descriptors DEI
- COINCIDENCE METHODS; FORECASTING; GE SEMICONDUCTOR DETECTORS; MONTE CARLO METHOD; RADIOISOTOPES; VALIDATION
- Descriptors DEC
- CALCULATION METHODS; COUNTING TECHNIQUES; ISOTOPES; MEASURING INSTRUMENTS; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS; TESTING
Optional Information
- Notes
- 11 refs.