Establishment of standard x-ray qualities to be used in diagnostic level at SSDLs

  • 1. Department of physics, Faculty of Science, University of Khartoum, Khartoum (Sudan)

Description

The objective of the present work is to modify x-ray machine (used by the secondary standard dosimetry laboratory in Sudan) to produce x-ray qualities suitable for the calibration at diagnostic level. It based on experimental evaluations. The importance of this study appears in development of protocols to calibrate the instruments used in diagnostic radiology in by good response of the instruments and low cost. In particular, the half value layer (HVL) values for the following qualities 40, 60, 80, 100, 120 and 150 were determined using various attenuation layers. Ionization chamber was used to determine the free air kerma rate at a distance of 100 cm from the x-ray generator. The obtained HVL values were compared with standard values of diagnostic levels. It was observed that the HVL of the present x-ray machine in most qualities are smaller than the standard ones. An approved method (described in the standard IEC 61267) was applied to determine the amount of additional filtration required to meet the standard values. The results show that the available cupper layers were not suitable for this purpose as most of the x-ray was absorbed in the material. Aluminum layers, on the other hand , show good performances to reduce the beams to the desired levels. The amount of additional filtration (from aluminum layers) needed in order to establish standard x-ray qualities to be used in diagnostic level were determined.(Author)

Availability note (English)

Available from INIS in electronic form; Also available from Sudan Atomic Energy Commission, Khartoum (SD)

Files

39088892.pdf

Files (703.1 kB)

Name Size Download all
md5:ea5023b947117e0d7fda46c973dc6970
703.1 kB Preview Download

Additional details

Publishing Information

Imprint Place
Khartoum (Sudan)
Imprint Pagination
50 p.
Report number
INIS-SD--322

INIS

Country of Publication
Sudan
Country of Input or Organization
Sudan
INIS RN
39088892
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Numerical Data, Thesis
Descriptors DEI
CALIBRATION; DIAGNOSTIC TECHNIQUES; EXPERIMENTAL DATA; IONIZATION CHAMBERS; KERMA; LEVELS; MEASURING INSTRUMENTS; RADIOLOGY; SSDL; SUDAN; X-RAY EQUIPMENT
Descriptors DEC
AFRICA; ARAB COUNTRIES; DATA; DEVELOPING COUNTRIES; EQUIPMENT; INFORMATION; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; NUMERICAL DATA; RADIATION DETECTORS

Optional Information

Notes
18 refs., 5 tabs., 23 figs.