Variation of the practical peak voltage with the sample rate for a mammography waveform generator

  • 1. Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)
  • 2. Universidade de Sao Paulo (USP), SP (Brazil). Inst. de Eletrotecnica e Energia
  • 3. Pontificia Univ. Catolica de Sao Paulo (PUC-SP), SP (Brazil). Dept. de Fisica

Description

The Practical Peak Voltage (PPV) was introduced by the IEC 61676 standard as an electrical parameter strongly related to the image contrast, which is determined from the X-ray tube voltage waveform. This quantity is used in diagnostic radiology reference systems quality control. Measurements have been made using a mammography equipment model Senographe 700T, manufactured by GE (General Electric). This equipment works in the voltage range from 22 to 35 kV and the tube has a molybdenum anode. Its internal voltage divider was used to do the acquisition of the voltage data of the X-ray tube. To acquire the waveforms of the equipment, it has been used an acquisition board model NI 5911 (National Instruments) and a LabVIEW routine that calculates all the voltage parameters associated with the X-ray tube. During the measurements, the sample rate was varied from 100 samples/s to 1 Msamples/s for a 24 kV voltage reference. Results showed a variation less than 0.05 kV for the PPV value. At the same time, it was evaluated the maximum value of the tube waveform (kVp - maximum). The results have shown that the value of kVp (maximum) varies up to 1 kV, for the higher sample rates, and have confirmed the low dependence of PPV with the sample rate of the tube waveform, in comparison with kVp (maximum), in mammography range. (author)

Part of:
Proceedings of the INAC 2007 International nuclear atlantic conference. Nuclear energy and energetic challenges for 21st. century. 15. Brazilian national meeting on reactor physics and thermal hydraulics; 8. Brazilian national meeting on nuclear applications

Additional details

Publishing Information

ISBN
978-85-99141-02-1
Imprint Title
Proceedings of the INAC 2007 International nuclear atlantic conference. Nuclear energy and energetic challenges for 21st. century. 15. Brazilian national meeting on reactor physics and thermal hydraulics; 8. Brazilian national meeting on nuclear applications
Imprint Pagination
[vp.]
Journal Page Range
[6 p.]

Conference

Title
INAC 2007 International nuclear atlantic conference. Nuclear energy and energetic challenges for 21st. century. 15. Brazilian national meeting on reactor physics and thermal hydraulics; 8. Brazilian national meeting on nuclear applications
Dates
30 Sep - 5 Oct 2007
Place
Santos, SP (Brazil)

INIS

Country of Publication
Brazil
Country of Input or Organization
Brazil
INIS RN
39086823
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference, Numerical Data, Non-conventional Literature
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; CALIBRATION; ELECTRIC POTENTIAL; EXPERIMENTAL DATA; L CODES; MAMMARY GLANDS; PEAKS; QUALITY CONTROL; WAVE FORMS; X-RAY TUBES
Descriptors DEC
BODY; COMPUTER CODES; CONTROL; DATA; DIAGNOSTIC TECHNIQUES; ELECTRON TUBES; EQUIPMENT; GLANDS; INFORMATION; MEDICINE; NUCLEAR MEDICINE; NUMERICAL DATA; ORGANS; RADIOLOGY; X-RAY EQUIPMENT

Optional Information

Notes
7 refs., 3 figs., 1 tab. Code: E02_1668.pdf