Published August 21, 2004 | Version v1
Journal article

Effect of area x-ray beam equalization on image quality and dose in digital mammography

  • 1. Department of Radiological Sciences, University of California, Irvine, CA 92697 (United States)

Description

In mammography, thick or dense breast regions persistently suffer from reduced contrast-to-noise ratio (CNR) because of degraded contrast from large scatter intensities and relatively high noise. Area x-ray beam equalization can improve image quality by increasing the x-ray exposure to under-penetrated regions without increasing the exposure to other breast regions. Optimal equalization parameters with respect to image quality and patient dose were determined through computer simulations and validated with experimental observations on a step phantom and an anthropomorphic breast phantom. Three parameters important in equalization digital mammography were considered: attenuator material (Z = 13-92), beam energy (22-34 kVp) and equalization level. A Mo/Mo digital mammography system was used for image acquisition. A prototype 16 x 16 piston driven equalization system was used for preparing patient-specific equalization masks. Simulation studies showed that a molybdenum attenuator and an equalization level of 20 were optimal for improving contrast, CNR and figure of merit (FOM = CNR2/dose). Experimental measurements using these parameters showed significant improvements in contrast, CNR and FOM. Moreover, equalized images of a breast phantom showed improved image quality. These results indicate that area beam equalization can improve image quality in digital mammography

Availability note (English)

Available online at http://stacks.iop.org/0031-9155/49/3539/pmb4_16_003.pdf or at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
49
Journal Issue
16
Journal Page Range
p. 3539-3557
ISSN
0031-9155
CODEN
PHMBA7

Conference

Title
World congress on medical physics and biomedical engineering
Acronym
WC2003
Dates
24-29 Aug 2003
Place
Sydney (Australia)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
36031472
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; COMPUTERIZED SIMULATION; IMAGES; MAMMARY GLANDS; MOLYBDENUM; NOISE; PHANTOMS; RADIATION DOSES; RESPIRATORS; SIGNAL-TO-NOISE RATIO
Descriptors DEC
BODY; DIAGNOSTIC TECHNIQUES; DOSES; ELEMENTS; GLANDS; MEDICINE; METALS; MOCKUP; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; REFRACTORY METALS; SIMULATION; STRUCTURAL MODELS; TRANSITION ELEMENTS