Adequacy of source to image receptor distance with chest postero-anterior projection in digital radiology system

  • 1. Dept. of Radiology, Samsung Medical Center, Seoul (Korea, Republic of)
  • 2. Dept. of Radiology, Science, Hanseo University, Seosan (Korea, Republic of)
  • 3. Dept. of Radiology, Hallym University Hospital, Chuncheon (Korea, Republic of)
  • 4. Dept. of Radiology, Science, Seonam University, Namwon (Korea, Republic of)

Description

The purpose of this study is to evaluate propriety of using SID 180cm at Chest PA examination and to find effect of geometrical cause to the image. XGEO-GC80, INNOVISION-SH, CXDI-40EG detector and a chest phantom designed self-production was used for this study. Images were acquired at SID 180cm with changing the factor OID as 0, 75 and 83mm and were analyzed by Centricity Radiography RA1000 PACS system. Statistical program was used the SPSS (Version 22.0, SPSS, Chicago, IL, USA), p-value(under 0.05) was considered to be statistically significant. In OID 0 mm was enlarged about 2.7⁓3.5 mm than the actual degree of the HS, BS of phantom in all equipments. Compared with the calculated magnification has been expanded 1.6⁓2.8% when viewed. The OID 75 mm with OID 83 mm was extended from the CS and BS 6⁓8 mm range. Compared to the calculated values, the measured values are expanded from 6.1 to 7.9%. CS and BS according to the OID change showed a statistically significant difference (p<0.05) among each group, the post-analysis only OID 0 mm group appeared as an independent group, 75 mm and 83 mm are separated in the same group It was. But had no statistically significant difference could change depending on the OID (p>0.05), post-mortem analysis showed, both in the same group. Heart sizes appears larger than actual size 6⁓8 mm at chest PA examination which is enlarged 6.1~7.9% more than the actual theoretical value. We can find magnification of the image because of the increase of the OID due to technical limitations between cover of standing detector and the image plate. so we suggest to have occurred between them when considering the need to adjust the equipment installed by the SID to match the characteristics of the equipment

Additional details

Publishing Information

Journal Title
Journal of Radiological Science and Technology
Journal Volume
39
Journal Issue
2
Series
16 refs, 2 figs, 3 tabs
Journal Page Range
p. 135-142
ISSN
2288-3509

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
47114296
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
CHEST; COMPUTER CALCULATIONS; DESIGN; EQUIPMENT; GEOMETRY; IMAGES; RADIOLOGY
Descriptors DEC
BODY; MATHEMATICS; MEDICINE; NUCLEAR MEDICINE