Description

In this study, we investigated the correlation between 2-dimensional (2D) and 3D gamma analysis using the new PTW OCTAVIUS 4D system for various parameters. For this study, we selected 150 clinically approved volumetric-modulated arc therapy (VMAT) plans of head and neck (50), thoracic (esophagus) (50), and pelvic (cervix) (50) sites. Individual verification plans were created and delivered to the OCTAVIUS 4D phantom. Measured and calculated dose distributions were compared using the 2D and 3D gamma analysis by global (maximum), local and selected (isocenter) dose methods. The average gamma passing rate for 2D global gamma analysis in coronal and sagittal plane was 94.81% ± 2.12% and 95.19% ± 1.76%, respectively, for commonly used 3-mm/3% criteria with 10% low-dose threshold. Correspondingly, for the same criteria, the average gamma passing rate for 3D planar global gamma analysis was 95.90% ± 1.57% and 95.61% ± 1.65%. The volumetric 3D gamma passing rate for 3-mm/3% (10% low-dose threshold) global gamma was 96.49% ± 1.49%. Applying stringent gamma criteria resulted in higher differences between 2D planar and 3D planar gamma analysis across all the global, local, and selected dose gamma evaluation methods. The average gamma passing rate for volumetric 3D gamma analysis was 1.49%, 1.36%, and 2.16% higher when compared with 2D planar analyses (coronal and sagittal combined average) for 3 mm/3% global, local, and selected dose gamma analysis, respectively. On the basis of the wide range of analysis and correlation study, we conclude that there is no assured correlation or notable pattern that could provide relation between planar 2D and volumetric 3D gamma analysis. Owing to higher passing rates, higher action limits can be set while performing 3D quality assurance. Site-wise action limits may be considered for patient-specific QA in VMAT

Availability note (English)

Available from http://dx.doi.org/10.1016/j.meddos.2014.05.002

Additional details

Identifiers

DOI
10.1016/j.meddos.2014.05.002;
PII
S0958-3947(14)00061-2;

Publishing Information

Journal Title
Medical Dosimetry
Journal Volume
39
Journal Issue
4
Journal Page Range
p. 300-308
ISSN
0958-3947

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
46126758
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
COMPARATIVE EVALUATIONS; DOSIMETRY; ESOPHAGUS; HEAD; NECK; PATIENTS; PHANTOMS; QUALITY ASSURANCE; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY; VERIFICATION
Descriptors DEC
BODY; DIGESTIVE SYSTEM; DOSES; EVALUATION; MEDICINE; MOCKUP; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; STRUCTURAL MODELS; THERAPY