High energy X-rays reconstruction in the entire radiotherapy energy range
Creators
Description
We have previously developed an original method of high energy X-ray spectral reconstruction from measured transmission data on a 12MV beam. This reconstruction is done by direct resolution of the matrix system A*F=T where T is the measured transmission vector, F is the fraction of the signal per unit energy produced by photons of a given energy in the unattenuated beam and A is the attenuation matrix. This technique is using spectral vectorial algebra. The resolution has been extended and tested in the entire radiotherapy energy range with photon beams from 4 to 25 MV nominal energies. Dosimetric quantities such as attenuation curves, mean energies and mean attenuation coefficients have been recalculated from our reconstructed spectra. These quantities have been computed for different materials
Additional details
Identifiers
- PII
- 0167814096805561;
Publishing Information
- Journal Title
- Radiotherapy and Oncology
- Journal Volume
- 37
- Journal Issue
- 3
- Journal Page Range
- p. S33
- ISSN
- 0167-8140
- CODEN
- RAONDT
INIS
- Country of Publication
- Ireland
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 34044598
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ATTENUATION; HARD X RADIATION; PHOTON BEAMS; RADIOTHERAPY; TRANSFER MATRIX METHOD; X-RAY SPECTRA
- Descriptors DEC
- BEAMS; CALCULATION METHODS; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEDICINE; NUCLEAR MEDICINE; RADIATIONS; RADIOLOGY; SPECTRA; THERAPY; X RADIATION