Study of influence of neutron field and γ-ray self-absorption on PGNAA measurement
- 1. College of Materials Science and Technology,Nanjing University of Aeronautics and Astronautics, Nanjing (China)
Description
In PGNAA measurement, the distributions of neutron field and γ-ray self-absorption in samples have an influence on the measurement results. A corresponding correction model with two effects into consideration and a theoretical calculation were carried out with an aqueous solution element detection setup. The theoretical calculation was applied to measurement results of detecting device by Monte Carlo simulation. After modification on the theoretical model, the results show that influence of neutron field and γ-ray self-absorption on PGNAA measurement decreases. (authors)
Additional details
Publishing Information
- Journal Title
- Atomic Energy Science and Technology
- Journal Volume
- 48
- Journal Issue
- suppl
- Journal Page Range
- p. 802-806
- ISSN
- 1000-6931
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 48072471
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- AQUEOUS SOLUTIONS; COMPUTERIZED SIMULATION; CORRECTIONS; DISTRIBUTION; GAMMA DETECTION; MODIFICATIONS; MONTE CARLO METHOD; NEUTRON ACTIVATION ANALYSIS; NEUTRONS; PROMPT GAMMA RADIATION; SELF-ABSORPTION
- Descriptors DEC
- ABSORPTION; ACTIVATION ANALYSIS; BARYONS; CALCULATION METHODS; CHEMICAL ANALYSIS; DETECTION; DISPERSIONS; ELECTROMAGNETIC RADIATION; ELEMENTARY PARTICLES; FERMIONS; GAMMA RADIATION; HADRONS; HOMOGENEOUS MIXTURES; IONIZING RADIATIONS; MIXTURES; NONDESTRUCTIVE ANALYSIS; NUCLEONS; RADIATION DETECTION; RADIATIONS; SIMULATION; SOLUTIONS; SORPTION
Optional Information
- Notes
- 4 figs., 1 tab., 9 refs.