Development of waste drum monitoring system for plutonium estimation
Creators
- 1. Radiation Safety Systems Division, Bhabha Atomic Research Centre, Mumbai (India)
Description
The screening based waste drum monitor system using neutron detectors was developed and evaluated. Present paper describes the studies carried out on the system evaluation for Plutonium Estimation. The system uses three BF3 neutron detectors with 3.8 cm HDPE moderator for two extreme detectors and with 7 cm HDPE moderator for middle detector arranged all three detectors in linear geometry. The middle detector shows less background and better sensitivity than of the two. Present system can detect 85 mg of plutonium at counting time of minimum 1000 second or more. With changing the detectors geometry of the system to triangular geometry and replacing all the detectors with 7 cm moderator systems, performance as well as MDL can be improved. (author)
Additional details
Publishing Information
- Journal Title
- Radiation Protection and Environment
- Journal Volume
- 37
- Journal Issue
- 3-4
- Journal Page Range
- p. 176-178
- CODEN
- RPREFM
INIS
- Country of Publication
- India
- Country of Input or Organization
- India
- INIS RN
- 46112219
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ALPHA DECAY; BF3 COUNTERS; DESIGN; PLUTONIUM; RADIATION DOSES; RADIOACTIVE WASTES
- Descriptors DEC
- ACTINIDES; DECAY; DOSES; ELEMENTS; MATERIALS; MEASURING INSTRUMENTS; METALS; NEUTRON DETECTORS; NUCLEAR DECAY; PROPORTIONAL COUNTERS; RADIATION DETECTORS; RADIOACTIVE MATERIALS; TRANSURANIUM ELEMENTS; WASTES
Optional Information
- Notes
- 3 refs., 5 figs., 2 tabs.