Published June 23, 2010 | Version v1
Journal article

The ESRF Miniature Pulsed Magnetic Field System

  • 1. ESRF, 6 rue Jules Horowitz, F-38043 Grenoble (France)

Description

We have developed a portable system to provide pulsed magnetic fields on the ESRF X-ray beamlines. The complete system consists of a power supply, liquid Helium and liquid Nitrogen dewars with a siphon each, control electronics and a double cryostat for separate coil and sample cooling. The liquid nitrogen cooled solenoids reach a maximum field of 30 Tesla for a total pulse duration of one milisecond. They are constructed for optimised cooling rate after the pulse to obtain a high duty cycle, the repetition rate is five pulses per minute at maximum field. The sample is cooled in an independent Helium flow cryostat which is inserted into the bore of the magnet. The flow cryostat has a temperature range from 5 to 250 Kelvin with a direct contact between the sample and Helium flow. This overview gives a general presentation of the system and we will show recent results.

Additional details

Identifiers

Publishing Information

Journal Title
AIP Conference Proceedings
Journal Volume
1234
Journal Issue
1
Journal Page Range
p. 107-110
ISSN
0094-243X
CODEN
APCPCS

Conference

Title
10. international conference on radiation instrumentation
Acronym
SRI 2009
Dates
27 Sep - 2 Oct 2009
Place
Melbourne (Australia)

Optional Information

Notes