Published July 1991 | Version v1
Miscellaneous

Detectors for LIDAR type Thomson scattering diagnostics

Creators

  • 1. Stuttgart Univ. (Germany). Inst. fuer Plasmaforschung

Description

A report on the capability of the microchannel plate photomultiplier type (ITT F4128) presently used at the JET LIDAR Thomson Scattering System is given. Detailed investigation on time response, low noise amplification, shutter ratio, gating behaviour, linear mode of operation and saturation pulse recovery carried out during the design phase for LIDAR are presented. New investigation with respect to dc-and gated operation showed no measurable changes in sensitivity of this MCP photomultiplier. Comparing this type of detector with other MCP photomultipliers and with streak cameras some detection schemes for future LIDAR type diagnostic are proposed. (author). 24 refs, 15 figs, 5 tabs

Part of:
LIDAR Thomson scattering

Additional details

Publishing Information

Imprint Title
LIDAR Thomson scattering
Imprint Pagination
214 p.
Journal Page Range
p. 169-191.
Report number
INIS-mf--13799

Conference

Title
IAEA technical committee meeting on LIDAR Thomson scattering.
Dates
8-10 Apr 1991.
Place
Abingdon (United Kingdom).

INIS

Country of Publication
International Atomic Energy Agency (IAEA)
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
25029917
Subject category
S70: PLASMA PHYSICS AND FUSION TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
IMAGE INTENSIFIERS; JET TOKAMAK; OPTICAL RADAR; PHOTOMULTIPLIERS; PLASMA DIAGNOSTICS; SPECIFICATIONS; THOMSON SCATTERING
Descriptors DEC
CLOSED PLASMA DEVICES; INELASTIC SCATTERING; MEASURING INSTRUMENTS; PHOTOTUBES; RADAR; RANGE FINDERS; SCATTERING; THERMONUCLEAR DEVICES; TOKAMAK DEVICES