Published January 1, 2018 | Version v1
Journal article

Dosimetry of Al2O3 optically stimulated luminescent dosimeter at high energy photons and electrons

  • 1. School of Health Sciences, Universiti Sains Malaysia, 16150 Kota Bharu, Kelantan (Malaysia)
  • 2. Medical Physics and Secondary Standard Dosimatry Laboratory, Malaysian Nuclear Agency, 43000 Bangi, Selangor (Malaysia)

Description

The linearity of Al2O3 OSL dosimeters (OSLD) were evaluated for dosimetry works in clinical photons and electrons. The measurements were made at a reference depth of Zref according to IAEA TRS 398:2000 codes of practice at 6 and 10 MV photons and 6 and 9 MeV electrons. The measured dose was compared to the thermoluminescence dosimeters (TLD) and ionization chamber commonly used for dosimetry works for higher energy photons and electrons. The results showed that the measured dose in OSL dosimeters were in good agreement with the reported by the ionization chamber in both high energy photons and electrons. A reproducibility test also reported excellent consistency of readings with the OSL at similar energy levels. The overall results confirmed the suitability of OSL dosimeters for dosimetry works involving high energy photons and electrons in radiotherapy. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1757-899X/298/1/012023

Additional details

Publishing Information

Journal Title
IOP Conference Series. Materials Science and Engineering (Online)
Journal Volume
298
Journal Issue
1
Journal Page Range
[4 p.]
ISSN
1757-899X

Conference

Title
International Nuclear Science, Technology and Engineering Conference 2017
Acronym
iNuSTEC2017
Dates
25-27 Sep 2017
Place