Published November 2018 | Version v1
Journal article

Validation of tissue phantom ratio obtained from percentage depth dose with TPR from direct measurement in a water phantom

Creators

  • 1. V.T.S.M. Peripheral Cancer Hospital, Gulbarga (India)

Description

This study to validate the TPR data derived from PDD with TPR from direct measurement in water phantom. From our study we found that average dose error is 1% for 6X, 10X, 15X, 6FFF, 10FFF. So, if TPR is not measured directly on machine it may induce 1% uncertainty in dose calculation

Additional details

Publishing Information

Journal Title
Journal of Medical Physics
Journal Volume
43
Journal Issue
suppl.1
Journal Page Range
p. 107
CODEN
JMPHFE

Conference

Title
39. annual conference of Association of Medical Physicists of India
Acronym
AMPICON 2018
Dates
2-4 Nov 2018
Place
Chennai (India)

INIS

Country of Publication
India
Country of Input or Organization
India
INIS RN
50065656
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ANIMAL TISSUES; DEPTH DOSE DISTRIBUTIONS; DOSE RATES; LINEAR ACCELERATORS; PHANTOMS; RADIATION DOSES; THREE-DIMENSIONAL CALCULATIONS; WATER
Descriptors DEC
ACCELERATORS; BODY; DOSES; HYDROGEN COMPOUNDS; MOCKUP; OXYGEN COMPOUNDS; RADIATION DOSE DISTRIBUTIONS; SPATIAL DOSE DISTRIBUTIONS; STRUCTURAL MODELS