Published November 2018
| Version v1
Journal article
Validation of tissue phantom ratio obtained from percentage depth dose with TPR from direct measurement in a water phantom
Description
This study to validate the TPR data derived from PDD with TPR from direct measurement in water phantom. From our study we found that average dose error is 1% for 6X, 10X, 15X, 6FFF, 10FFF. So, if TPR is not measured directly on machine it may induce 1% uncertainty in dose calculation
Additional details
Publishing Information
- Journal Title
- Journal of Medical Physics
- Journal Volume
- 43
- Journal Issue
- suppl.1
- Journal Page Range
- p. 107
- CODEN
- JMPHFE
Conference
- Title
- 39. annual conference of Association of Medical Physicists of India
- Acronym
- AMPICON 2018
- Dates
- 2-4 Nov 2018
- Place
- Chennai (India)
INIS
- Country of Publication
- India
- Country of Input or Organization
- India
- INIS RN
- 50065656
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ANIMAL TISSUES; DEPTH DOSE DISTRIBUTIONS; DOSE RATES; LINEAR ACCELERATORS; PHANTOMS; RADIATION DOSES; THREE-DIMENSIONAL CALCULATIONS; WATER
- Descriptors DEC
- ACCELERATORS; BODY; DOSES; HYDROGEN COMPOUNDS; MOCKUP; OXYGEN COMPOUNDS; RADIATION DOSE DISTRIBUTIONS; SPATIAL DOSE DISTRIBUTIONS; STRUCTURAL MODELS