2D sparse array transducer optimization for 3D ultrasound imaging

  • 1. Hanyang University, Seoul (Korea, Republic of)

Description

A 3D ultrasound image is desired in many medical examinations. However, the implementation of a 2D array, which is needed for a 3D image, is challenging with respect to fabrication, interconnection and cabling. A 2D sparse array, which needs fewer elements than a dense array, is a realistic way to achieve 3D images. Because the number of ways the elements can be placed in an array is extremely large, a method for optimizing the array configuration is needed. Previous research placed the target point far from the transducer array, making it impossible to optimize the array in the operating range. In our study, we focused on optimizing a 2D sparse array transducer for 3D imaging by using a simulated annealing method. We compared the far-field optimization method with the near-field optimization method by analyzing a point-spread function (PSF). The resolution of the optimized sparse array is comparable to that of the dense array.

Additional details

Publishing Information

Journal Title
Journal of the Korean Society for Nondestructive Testing
Journal Volume
34
Journal Issue
6
Series
7 refs, 6 figs, 3 tabs
Journal Page Range
p. 441-446
ISSN
1225-7842

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
46084376
Subject category
S42: ENGINEERING;
Descriptors DEI
FABRICATION; MEDICAL EXAMINATIONS; OPTIMIZATION; RESOLUTION; TRANSDUCERS; ULTRASONIC TESTING
Descriptors DEC
ACOUSTIC TESTING; MATERIALS TESTING; MEDICAL SURVEILLANCE; NONDESTRUCTIVE TESTING; TESTING