Published August 1982 | Version v1
Journal article

Evaluation of a new distal delivery guidewire for steel coils

  • 1. Amsterdam Univ. (Netherlands). Dept. of Diagnostic Radiology, Pathology and Gastroenterology

Description

A new distal delivery guidewire (DDG) for steel coils is described. By means of a DDG 'mini' coils as well as 'new' coils can be delivered intravascularly through catheters with an inside diameter large enough to let pass a 0.97 mm. guidewire. Delivery using a DDG is compared with delivery using a pusher guidewire (conventional method); the former seems to be safer. Additionally, coil placement beyond the location of the catheter tip is only possible when using a DDG. Forty-nine venous and four arterial coil placements resulted in only one misplacement which could be corrected angiographically. (orig.)

Additional details

Publishing Information

Journal Title
Eur. J. Radiol.
Journal Volume
2
Journal Issue
3
Series
Eur. J. Radiol.
Journal Page Range
250-253
ISSN
0720-048X

INIS

Country of Publication
Netherlands
Country of Input or Organization
Germany
INIS RN
14724301
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; DIAGNOSIS; IMAGES; PATIENTS; TESTES; THERAPY; UROGENITAL SYSTEM DISEASES; VEINS
Descriptors DEC
BLOOD VESSELS; BODY; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; DISEASES; GONADS; MALE GENITALS; MEDICINE; ORGANS