Evaluation of a mammographic stereotactic device for localization, fine-needle aspiration cytologic studies, and core biopsy of suspicious lesions

Description

Mammography-guided interventional breast procedures, such as preoperative localization, fine-needle aspiration cytology (FNAC), and core biopsy of suspicious lesions, require accurate three-dimensional localization. The authors have evaluated a prototype stereotactic mammography device for localizing abnormalities with both phantom and clinical studies. Twenty-six localizations on a phantom were within 0.5 mm +- 0.93 (standard deviation) from the lesions; accuracy in clinical procedures was within 0.6 mm +- 0.8. Procedures are performed an average of 8 minutes faster with this device. They are prospectively evaluating mammography-guided FNAC and core biopsy of suspicious lesions in 100 patients. Results of FNAC performed without the stereotactic device agreed with results of open surgical biopsy in six of eight patients; results of core biopsy agreed in seven of eight

Additional details

Publishing Information

Publisher
Radiological Society of North America Inc.
Imprint Place
Oak Brook, IL (USA)
Imprint Title
Radiological Society of North America 74th scientific assembly and annual meeting (Abstracts)
Imprint Pagination
395 p.
Journal Page Range
p. 351.

Conference

Title
74. scientific assembly and annual meeting of the Radiological Society of North America (RSNA).
Dates
27 Nov - 2 Dec 1988.
Place
Chicago, IL (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
21002803
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ACCURACY; BIOMEDICAL RADIOGRAPHY; BIOPSY; CYTOLOGICAL TECHNIQUES; DIAGNOSIS; EQUIPMENT; EVALUATION; FEASIBILITY STUDIES; MAMMARY GLANDS; PATHOLOGICAL CHANGES; PATTERN RECOGNITION; SURGERY
Descriptors DEC
BODY; DIAGNOSTIC TECHNIQUES; DISEASES; GLANDS; MEDICINE; ORGANS

Optional Information

Secondary number(s)
CONF-8811134--.