A study on error characteristics of large size electromagnetic flowmeter in the range of low velocity
Creators
- 1. Korea Water Resources Corp., Daejeon (Korea, Republic of)
- 2. Chungnam National Univ., Daejeon (Korea, Republic of)
Description
The large size electromagnetic flowmeter was tested to investigate the variation of its error characteristics in the range of low velocity under 0.6 m/s using flowmeter calibration system. For the two case of valve opening rate 100 % and 50 %, these tests were undertaken three times each for twelve velocity condition from 0.05 m/s to 0.6 m/s with increment of 0.05 m/s. It is shown that error characteristic of electromagnetic flowmeter was stabilized within ±0.4 % of rate both higher than 0.25 m/s of velocity condition and 50 % of valve opening position. But, measurement deviation of flowmeter for φ400 mm and φ600 mm was out of expected deviation range. It is necessary to correction with calibration. In conclusion, error characteristic of electromagnetic flowmeter wasn't changed proportion to its size
Additional details
Publishing Information
- Journal Title
- Transactions of the Korean Society of Mechanical Engineers. B
- Journal Volume
- 32
- Journal Issue
- 3
- Series
- 7 refs, 7 figs, 1 tab
- Journal Page Range
- p. 235-240
- ISSN
- 1226-4881
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 40052524
- Subject category
- S42: ENGINEERING;
- Descriptors DEI
- BENCH-SCALE EXPERIMENTS; CALIBRATION; ELECTROMECHANICS; ERRORS; FLOWMETERS; VALVES
- Descriptors DEC
- CONTROL EQUIPMENT; EQUIPMENT; FLOW REGULATORS; MEASURING INSTRUMENTS; MECHANICS; METERS