Published October 1980 | Version v1
Report Restricted

Considerations for design parameters for a dedicated medical accelerator

Description

There are only a very few critical parameters which determine the size, performance and cost of a heavy ion accelerator. These are the mass of the heaviest ion desired, the maximum range of this heaviest ion in tissue, and the highest intensity desired. Other parameters, such as beam emittance, beam delivery flexibility, reliability and experimental facility configurations are important, but are not primary driving factors in the design effort. The various clinical applications for a heavy ion accelerator are evaluated, detailing the most desirable beams for each application

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS., PC A03/MF A01.

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Additional details

Publishing Information

Imprint Pagination
27 p.
Report number
LBL--11685

Conference

Title
MARIA workshop II on optimal parameters of relativistic heavy ions for radio-therapy research.
Dates
9 - 10 Oct 1980.
Place
Edmonton, Alberta, Canada.

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
12596351
Subject category
S43: PARTICLE ACCELERATORS;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ACCELERATOR FACILITIES; HEAVY ION ACCELERATORS; NUCLEAR MEDICINE; RADIOTHERAPY; SPECIFICATIONS; USES
Descriptors DEC
ACCELERATORS; MEDICINE; THERAPY

Optional Information

Secondary number(s)
CONF-8010145--1.