Published March 2015 | Version v1
Journal article

A method for modeling laterally asymmetric proton beamlets resulting from collimation

  • 1. Department of Radiation Oncology, University of Iowa, 200 Hawkins Drive, Iowa City, Iowa 52242 (United States)
  • 2. Department of Human Oncology, University of Wisconsin, 600 Highland Avenue, Madison, Wisconsin 53792 (United States)
  • 3. Division of Medical Physics, CDH Proton Center, 4455 Weaver Parkway, Warrenville, Illinois 60555 (United States)

Description

Purpose: To introduce a method to model the 3D dose distribution of laterally asymmetric proton beamlets resulting from collimation. The model enables rapid beamlet calculation for spot scanning (SS) delivery using a novel penumbra-reducing dynamic collimation system (DCS) with two pairs of trimmers oriented perpendicular to each other. Methods: Trimmed beamlet dose distributions in water were simulated with MCNPX and the collimating effects noted in the simulations were validated by experimental measurement. The simulated beamlets were modeled analytically using integral depth dose curves along with an asymmetric Gaussian function to represent fluence in the beam’s eye view (BEV). The BEV parameters consisted of Gaussian standard deviations (sigmas) along each primary axis (σx1x2y1y2) together with the spatial location of the maximum dose (μxy). Percent depth dose variation with trimmer position was accounted for with a depth-dependent correction function. Beamlet growth with depth was accounted for by combining the in-air divergence with Hong’s fit of the Highland approximation along each axis in the BEV. Results: The beamlet model showed excellent agreement with the Monte Carlo simulation data used as a benchmark. The overall passing rate for a 3D gamma test with 3%/3 mm passing criteria was 96.1% between the analytical model and Monte Carlo data in an example treatment plan. Conclusions: The analytical model is capable of accurately representing individual asymmetric beamlets resulting from use of the DCS. This method enables integration of the DCS into a treatment planning system to perform dose computation in patient datasets. The method could be generalized for use with any SS collimation system in which blades, leaves, or trimmers are used to laterally sharpen beamlets

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
42
Journal Issue
3
Journal Page Range
p. 1321-1321.14
ISSN
0094-2405
CODEN
MPHYA6

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
46119338
Subject category
S07: ISOTOPES AND RADIATION SOURCES;
Descriptors DEI
ASYMMETRY; BEAM PROFILES; COMPUTERIZED SIMULATION; DEPTH DOSE DISTRIBUTIONS; MONTE CARLO METHOD; PROTON BEAMS; RADIOTHERAPY
Descriptors DEC
BEAMS; CALCULATION METHODS; MEDICINE; NUCLEAR MEDICINE; NUCLEON BEAMS; PARTICLE BEAMS; RADIATION DOSE DISTRIBUTIONS; RADIOLOGY; SIMULATION; SPATIAL DOSE DISTRIBUTIONS; THERAPY

Optional Information

Notes
(c) 2015 American Association of Physicists in Medicine