Published June 2008 | Version v1
Miscellaneous

Development of a high pressure Xe ionization chamber for environmental radiation spectroscopy

  • 1. Korea Atomic Energy Research Inst., Daejeon (Korea, Republic of)
  • 2. Dept. of Nuclear Engineering, Hanyang Univ., Seoul (Korea, Republic of)
  • 3. Dept. of Environmental Engineering, Yonsei Univ., Wonju (Korea, Republic of)

Description

A High Pressure Xenon ionization chamber is a promising radiation detector for environmental radiation measurement due to its radiation hardness, its physical rigidity, and its capability of operation at a high temperature up to about 170degC. A cylindrical high pressure xenon ionization chamber, which was configured with a shielding mesh to improve its energy resolution, was designed on the basis of an electron transfer simulation code (EGSnrc) to extract an optimal density of Xe gas and a thickness of the chamber wall. An electron drift simulation code, Garfield, which was coupled with a Maxwell electric filed calculator, was also employed for the electron drift simulations due to the geometry of the shielding mesh. Shielding inefficiency was also calculated. A spherical ionization chamber was also designed and fabricated to monitor environmental radiation. A noble gas system was constructed to create a noble gas with a high purity and to inject the noble gas up to 60 atm. The combination of an oxygen absorbent (Oxisorb), a molecular sieve, and a high temperature getter can minimize the electro-negative impurities, such as the O2 and N2 gas, to below about several ppb levels. Preliminary tests such as leakage currents, saturation currents, and gas leak test were performed. The performance of the two fabricated ionization chambers at a low dose rate was tested by using a conventional shadow technique with a NIST certified 33.52 MBq 226Ra source in the calibration room at KAERI. (author)

Part of:
Proceedings of the fourth international symposium on radiation safety and detection technology

Additional details

Publishing Information

Imprint Title
Proceedings of the fourth international symposium on radiation safety and detection technology
Imprint Pagination
765 p.
Journal Page Range
p. 383-386

Conference

Title
4. international symposium on radiation safety and detection technology
Acronym
ISORD-4
Dates
18-20 Jul 2007
Place
Seoul (Korea, Republic of)

INIS

Country of Publication
Japan
Country of Input or Organization
Japan
INIS RN
51059346
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
CURRENTS; DOSE RATES; ELECTRIC FIELDS; ELECTRON DRIFT; ELECTRON TRANSFER; ENERGY RESOLUTION; FANO FACTOR; IMPURITIES; IONIZATION; IONIZATION CHAMBERS; RADIATION MONITORING; SHIELDING; XENON
Descriptors DEC
DIMENSIONLESS NUMBERS; ELEMENTS; FLUIDS; GASES; MEASURING INSTRUMENTS; MONITORING; NONMETALS; RADIATION DETECTORS; RARE GASES; RESOLUTION

Optional Information

Notes
8 refs., 8 figs., 1 tab.; This record replaces 43053542 Imprint:Published as Supplement 5 (Jun 2008) of Journal of Nuclear Science and Technology. Publication month in colophon is Aug 2008