Published September 10, 1986 | Version v1
Journal article

Performance of a segmented Ge/BGO detector as an escape suppressed and pair spectrometer

  • 1. Gesellschaft fuer Schwerionenforschung m.b.H., Darmstadt (Germany, F.R.)

Description

A γ-spectrometer consisting of one planar and two large volume coaxial germanium crystals, closely packed in a common cryostat, and of a cylindrically symmetric bismuth germanate shield has been built. The system can simultaneously be operated in various escape suppressed modes as well as in a pair spectrometer mode, choosing appropriate coincidence and anticoincidence conditions. This unique feature of the system gives rise to both an exceptional wide dynamical range in γ-ray energies and a good background suppression. The performance of the system has been measured in an energy range between 120 keV and 3.4 MeV using standard radioactive sources. (orig.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods Phys. Res., Sect. A
Journal Volume
249
Journal Issue
2/3
Series
Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
320-324
ISSN
0168-9002
CODEN
NIMAE

Optional Information

Notes
Also published as report GSI--86-6(prepr.).