Published December 2020 | Version v1
Journal article

A Study on the Dose Assessment of Nanotungsten Shielding Body in Children's Abdominal and Pelvic X-ray Examination

  • 1. Kyung Hee University Hostpital at Gangdong, Seoul (Korea, Republic of)
  • 2. Donggang University, Gwangju (Korea, Republic of)

Description

A shield made of nano-sized tungsten particles was fabricated using eco-friendly materials to protect organs and gonads in the abdomen of children who are sensitive to abdominal and pelvic X-ray examination. We would like to evaluate its usefulness through dose evaluation. During the abdominal and pelvic X-ray examination of children, there was a dose reduction of about 60% in ESD values before and after the use of the shield. If the Nano Tungsten shield is commercialized and used, it is believed that exposure doses for pediatric patients' radiation-related tests such as general X-ray imaging, CT examination, and fluoroscopic examination can be reduced

Additional details

Publishing Information

Journal Title
Journal of Radiation Industry
Journal Volume
14
Journal Issue
4
Series
16 refs, 9 figs, 7 tabs
Journal Page Range
p. 399-405
ISSN