Published 1992 | Version v1
Book

Estimation of plutonium in presence of uranium by ICP-AES

  • 1. Bhabha Atomic Research Centre, Bombay (India). Fuel Reprocessing Div.

Description

An ICP spectrometric method has been standardized for the estimation of plutonium in presence of uranium in the highly radioactive Purex process solutions without any separation or pre-treatment but with the appropriate dilutions of the sample to reduce the radiation dose-rate to an acceptable level for safe handling in the glove box. The detection limit for Pu is 50 μg/l and the precision is ±5%. (author). 4 refs., 2 tabs

Part of:
Nuclear and radiochemistry symposium [Preprint volume]

Additional details

Publishing Information

Publisher
Department of Atomic Energy.
Imprint Place
Bombay (India)
Imprint Title
Nuclear and radiochemistry symposium [Preprint volume]
Imprint Pagination
482 p.
Journal Page Range
p. 357-358.

Conference

Title
Nuclear and radiochemistry symposium.
Dates
21-24 Dec 1992.
Place
Visakhapatnam (India).

Optional Information