Published April 2000
| Version v1
Journal article
Simulation and measurement of air generated electron contamination in radiotherapy
Description
Monte Carlo simulations using MCNP4A and experimental work have been performed to evaluate the contribution of electrons that are produced in air from a 6 MV x-ray beam produced by a medical linear accelerator. Results show that up to 9% of applied dose is delivered to a patient's skin surface from electrons excited in the irradiated air column. For most field sizes this constitutes approximately 30% of dose delivered to this region. This study gives clinically relevant information on the origins and quantity of electrons at the skin of patients undergoing radiation therapy
Additional details
Identifiers
- PII
- S1350448799002632;
Publishing Information
- Journal Title
- Radiation Measurements
- Journal Volume
- 32
- Journal Issue
- 2
- Journal Page Range
- p. 105-111
- ISSN
- 1350-4487
- CODEN
- RMEAEP
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- Turkey
- INIS RN
- 33048510
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- CONTAMINATION; ELECTRONS; LINEAR ACCELERATORS; MATHEMATICS; MEASURE THEORY; MEDICAL EXAMINATIONS; RADIATION DOSES; RADIOTHERAPY; SIMULATION; SKIN
- Descriptors DEC
- ACCELERATORS; BODY; DOSES; ELEMENTARY PARTICLES; FERMIONS; LEPTONS; MATHEMATICS; MEDICAL SURVEILLANCE; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; THERAPY
Optional Information
- Copyright
- Copyright (c) 2000 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.