Published January 1986 | Version v1
Journal article

Variation of gamma camera photographic image density with count rate

  • 1. SUNY Upstate Medical Center, Syracuse

Description

In recent years there has been a gradual shift in gamma camera image recording toward the use of transparency films. When this occurs one encounters the intermittency effect. This effect causes the image density to increase as the count rate increases. In addition, density may also increase because the light output per count from the cathode ray tube also increases as the count rate increases. If these effects are not taken into account, images obtained at low count rates may be too light and those obtained at high count rates may be too dark. This paper discusses these phenomena and presents experimental data that shows how image density varies with respect to count rate for several commonly used films

Additional details

Publishing Information

Journal Title
J. Nucl. Med.
Journal Issue
no.1
Series
J. Nucl. Med.
ISSN
0022-3123
CODEN
JNMEA

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
17068518
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; GAMMA CAMERAS; IMAGES; PHOTOGRAPHIC FILMS; SCINTILLATION COUNTING; VARIATIONS
Descriptors DEC
CAMERAS; COUNTING TECHNIQUES; DIAGNOSTIC TECHNIQUES; MEDICINE